Zone III REBOA and the COBRA-OS™: Safety of Inadvertent Iliac Artery Device Deployment

Authors

  • Adam Power Western University
  • Asha Parekh Western University
  • Tyler Beveridge Western University
  • Adam Groh Western University
  • Laura J Moore University of Texas McGovern Medical School

DOI:

https://doi.org/10.26676/jevtm.v5i2.202

Keywords:

REBOA, Iliac, Safety

Abstract

Background: REBOA is an emerging technique in trauma. However, inadvertent iliac artery balloon inflation can lead to complications. This study aims to investigate the safety characteristics of the COBRA-OS™ compared to a 7 Fr commercially available device during purposeful iliac artery balloon overinflation.

Methods:  In vitro: the COBRA-OS™ was inflated in explanted porcine iliac arteries and intentionally overinflated until balloon or vessel rupture occurred. In vivo: the COBRA-OS™ and 7 Fr device were deployed in the iliac arteries and intentionally overinflated until rupture of the balloon or blood vessel.

Results: In vitro: an average volume of 1mL was required for occlusion using the COBRA-OS™ and the mean balloon rupture volume was 32.5mL. The COBRA-OS™ partially migrated into the aorta in all cases. In vivo: the COBRA-OS™ and 7 Fr device occluded the iliac arteries with a mean volume of 3.5 mL.  Overinflation resulted in no iliac ruptures with the COBRA-OS™ (mean balloon rupture volume = 10mL). Overinflation with the 7 Fr device resulted in 1 iliac rupture at 5mL. The other two 7 Fr devices had a mean balloon rupture volume of 5 mL. All COBRA-OS™ devices moved partially up into the aorta during inflation while all 7 Fr devices remained in the iliac artery.

Conclusions: The COBRA-OS™ allows for significant overinflation when deployed in the common iliac artery of a porcine model due to its unique design. This ultimately may help to prevent balloon and blood vessel rupture during clinical use, however further studies are required.

Downloads

Additional Files

Published

2021-10-03

Issue

Section

Original Article